How is protein concentration measured?

 

The standard procedure for measuring protein concentration is UV analysis, wherein in the protein concentration is calculated from the measured absorbance at a given wavelength, using the Beer-Lambert expression shown below, where A is the absorbance, e is the molar absorptivity, b is the path length, and C is the molar concentration.

 

image\img00118.gif

 

A common mistake that is made in the use of Beer’s Law to determine protein concentration, is failure to recognize that the expression is only valid within a narrow range of protein concentrations. Figure 1 shows the concentration dependence of the UV spectra for lysozyme in PBS (pH 7.6). The dashed line indicates the wavelength (281.5 nm) for the lysozyme absorptivity of 2.64 L cm-1 g-1 reported in the Worthington Enzyme Manual.

 

image\img00119.gif

 

Figure 1: Concentration dependence of the UV spectra for lysozyme in PBS, measured in a 1 cm path length cell.

 

The concentration dependence of the absorbance at 281.5 nm for the lysozyme series is shown in Figure 2. As seen in this figure, a negative deviation from Beer’s law is observed for concentrations > 1.0 g/L. When determining the concentration of an unknown lysozyme sample then, one would need to insure that the measurement was conducted under conditions wherein the concentration was < 1 g/L.

 

image\img00120.gif

 

Figure 2: Beer’s Law plot for lysozyme in PBS, measured at a wavelength of 281.5 nm.

 

The upper concentration limit for Beer’s Law determination of protein concentration is unique to each protein. Consider for example, the Beer’s Law plot shown in Figure 3 for BSA in 0.l M NaCl at pH 4.0. In contrast to lysozyme with an upper concentration limit of 1 mg/mL, BSA absorbance is linear with concentration up to circa 4 mg/mL.

 

image\img00121.gif

 

Figure 3: Beer’s Law plot for BSA in 0.1 M NaCl & pH 4.0, measured at 279 nm.

 

The ideal conditions for protein concentration determinations using the Beer-Lambert expression can be identified by conducting an analysis of error propagation. In the absence of scattering, the wavelength dependent absorbance (A) and transmittance (T) of a solution of particles are related as shown below, where I is the sample intensity, Io is the incident intensity, e is the absorptivity, b is the cell path length, and C is the analyte concentration.

 

image\img00122.gif

 

After rearrangement, the differential of C with respect to T can be written as follows.

 

image\img00123.gif

 

Assuming dT » DT, the relative error in the concentration arising from instrumental error in the transmittance reading can then be approximated by the following expression.

 

image\img00124.gif

 

The relative error in the concentration as a function of %T and absorbance for arbitrarily selected DT values of 0.2 and 1.0 % is shown in Figures 4a & b. The relative error in concentration displays a minimum at 38.5 %Transmittance, independent of the magnitude of instrumental error. The corresponding optimum absorbance value is 0.415. Hence, when determining unknown sample concentrations via UV measurements, the effects of instrumental error can be minimized by incrementally diluting the samples to a UV absorbance value of circa 0.4.

 

image\img00125.gif

 

Figure 4a: The relative error in concentration as a function of %T for instrumental error (DT) values of 0.2 & 1.0 %T.

 

image\img00126.gif

 

Figure 4b: The relative error in concentration as a function of the absorbance for instrumental error (DT) values of 0.2 & 1.0 %T.

 

For lysozyme (e = 2.64 L cm-1 g-1) and BSA (e = 0.62 L cm-1 g-1), this "ideal" absorbance of 0.415 corresponds to concentrations of 0.16 and 0.67 mg/mL respectively, if the measurements are collected in a 1 cm path length cell. If a shorter path length cell, such as a 2 mm cell is used, the optimal concentrations for UV measurements of lysozyme and BSA would be increased to 0.79 and 3.35 mg/mL respectively.

 

 

For additional questions or information regarding Malvern Instruments complete line of particle and materials characterization products, visit us at www.malvern.com.

 